Overnight Closed-Loop Control Improves Glycemic Control in a Multicenter Study of Adults With Type 1 Diabetes.

Citation
Brown, S. A., et al. “Overnight Closed-Loop Control Improves Glycemic Control In A Multicenter Study Of Adults With Type 1 Diabetes.”. The Journal Of Clinical Endocrinology And Metabolism, pp. 3674-3682.
Center Albert Einstein College of Medicine
Author Sue A Brown, Marc D Breton, Stacey M Anderson, Laura Kollar, Patrick Keith-Hynes, Carol J Levy, David W Lam, Camilla Levister, Nihat Baysal, Yogish C Kudva, Ananda Basu, Vikash Dadlani, Ling Hinshaw, Shelly McCrady-Spitzer, Daniela Bruttomesso, Roberto Visentin, Silvia Galasso, Simone Del Favero, Yenny Leal, Federico Boscari, Angelo Avogaro, Claudio Cobelli, Boris P Kovatchev
Abstract

Context: Closed-loop control (CLC) for the management of type 1 diabetes (T1D) is a novel method for optimizing glucose control, and strategies for individualized implementation are being developed.

Objective: To analyze glycemic control in an overnight CLC system designed to "reset" the patient to near-normal glycemic targets every morning.

Design: Randomized, crossover, multicenter clinical trial.

Participants: Forty-four subjects with T1D requiring insulin pump therapy.

Intervention: Sensor-augmented pump therapy (SAP) at home vs 5 nights of CLC (active from 23:00 to 07:00) in a supervised outpatient setting (research house or hotel), with a substudy of 5 nights of CLC subsequently at home.

Main Outcome Measure: The percentage of time spent in the target range (70 to 180 mg/dL measured using a continuous glucose monitor).

Results: Forty subjects (age, 45.5 ± 9.5 years; hemoglobin A1c, 7.4% ± 0.8%) completed the study. The time in the target range (70 to 180 mg/dL) significantly improved in CLC vs SAP over 24 hours (78.3% vs 71.4%; P = 0.003) and overnight (85.7% vs 67.6%; P < 0.001). The time spent in a hypoglycemic range (<70 mg/dL) decreased significantly in the CLC vs SAP group over 24 hours (2.5% vs 4.3%; P = 0.002) and overnight (0.9% vs 3.2%; P < 0.001). The mean glucose level at 07:00 was lower with CLC than with SAP (123.7 vs 145.3 mg/dL; P < 0.001). The substudy at home, involving 10 T1D subjects, showed similar trends with an increased time in target (70 to 180 mg/dL) overnight (75.2% vs 62.2%; P = 0.07) and decreased time spent in the hypoglycemic range (<70 mg/dL) overnight in CLC vs SAP (0.6% vs 3.7%; P = 0.03).

Conclusion: Overnight-only CLC increased the time in the target range over 24 hours and decreased the time in hypoglycemic range over 24 hours in a supervised outpatient setting. A pilot extension study at home showed a similar nonsignificant trend.

Year of Publication
2017
Journal
The Journal of clinical endocrinology and metabolism
Volume
102
Issue
10
Number of Pages
3674-3682
Date Published
12/2017
ISSN Number
1945-7197
DOI
10.1210/jc.2017-00556
Alternate Journal
J. Clin. Endocrinol. Metab.
PMID
28666360
PMCID
PMC5630248